Perbedaan Hasil Pemeriksaan Kolesterol Total dalam Darah Menggunakan Alat Architect C4000 dan Alinity C1000
DOI:
https://doi.org/10.61132/obat.v3i6.1802Keywords:
Alinity, Architect, Cholesterol, Enzymatic colorimetric, HypercholesterolemiaAbstract
Total cholesterol measurement is essential for assessing cardiovascular risk, but consistency between laboratory instruments must be validated for reliable results. Architect C4000 and Alinity C1000 operate on enzymatic principles but differ in system design, reagents, and supporting technology. This study aimed to compare total cholesterol measurements between the two instruments using a non-parametric statistical approach (Mann-Whitney test). A comparative analytical design was conducted with 100 serum samples obtained via purposive sampling and divided for measurement with each device. Statistical analysis evaluated differences, and mean differences were calculated using the Hodges-Lehman method. Results showed mean cholesterol values of 182.98 mg/dL for Architect and 182.37 mg/dL for Alinity, with no significant difference (p=0.9942) and a median difference of 0.0 (95% CI: -9.0 to 9.0). Data distribution was nearly identical for both instruments in terms of mean and spread. Therefore, both methods demonstrate comparable validity in total cholesterol measurement, as they rely on similar principles and technologies, with reagents calibrated to the same standards.
Downloads
References
Aisya, R. W., & Dharmawati, L. (2021). Hubungan kebiasaan konsumsi makanan cepat saji dengan kejadian penyakit jantung koroner pada pasien rawat jalan di RSUD Dr. Moewardi. Jurnal Medika Indonesia, 2(2), 21–28.
Bishop, M. L. (2020). Clinical chemistry: Principles, techniques, and correlations, enhanced edition. Jones & Bartlett Learning.
Caeiro, A., Caeiro, S., Correia, S., & Canhoto, J. (2022). Induction of somatic embryogenesis in tamarillo (Solanum betaceum Cav.) involves increases in the endogenous auxin indole-3-acetic acid. Plants, 11(10), 1347. https://doi.org/10.3390/plants11101347
Desoky, E.-S. M., Mansour, E., Ali, M. M. A., Yasin, M. A. T., Abdul-Hamid, M. I. E., Rady, M. M., & Ali, E. F. (2021). Exogenously used 24-epibrassinolide promotes drought tolerance in maize hybrids by improving plant and water productivity in an arid environment. Plants, 10(2), 354. https://doi.org/10.3390/plants10020354
Douxfils, J., Morimont, L., Delvigne, A.-S., Devel, P., Masereel, B., Haguet, H., Bouvy, C., & Dogné, J.-M. (2020). Validation and standardization of the ETP-based activated protein C resistance test for the clinical investigation of steroid contraceptives in women: An unmet clinical and regulatory need. Clinical Chemistry and Laboratory Medicine, 58(2), 294–305. https://doi.org/10.1515/cclm-2019-0471
Fan, G., & Wang, Q. (2021). Quality control and quality assurance. In Clinical molecular diagnostics (pp. 97–113). Springer. https://doi.org/10.1007/978-981-16-1037-0_8
Juraijin, D. (2025). Quality assurance analysis internal analytical phase of urine microalbumin examination in patients hemodialisa. Journal Health Applied Science and Technology, 3(1), 25–30. https://doi.org/10.52523/jhast.v3i1.68
Kopacz, A., Kubicka-Russel, D., Liszewski, G., Bukowska, A., Samek, S., Malka, D., Łętowska, M., & Grabarczyk, P. (2024). Evaluation and experience from routine use of chemiluminescence assays for serological screening of blood and plasma donations on the Alinity s system and the Alinity i system, two new fully automated immunoassay systems in Poland. Practical Laboratory Medicine, 39, e00364. https://doi.org/10.1016/j.plabm.2024.e00364
Lapić, I., Šegulja, D., Dukić, K., Bogić, A., Lončar Vrančić, A., Komljenović, S., Šparakl, T., Grdiša Teodorović, K., Cigula Kurajica, V., & Baršić Lapić, I. (2022). Analytical validation of 39 clinical chemistry tests and 17 immunoassays on the Alinity analytical system. Scandinavian Journal of Clinical and Laboratory Investigation, 82(3), 199–209. https://doi.org/10.1080/00365513.2022.2056856
Liana, P., Jenica, A., Suciati, T., Rahmawati, E., Pariyana, P., & Umar, T. P. (2023). Comparison of liver function test results between Architect C8000 and COBAS C501 automatic chemistry analyzer. Archives of Razi Institute, 78(3), 1141.
Moonti, M. A., Sutandi, A., & Fitriani, N. D. (2023). Hubungan life style dengan kejadian hipertensi pada dewasa di Desa Jagara Kecamatan Darma Kabupaten Kuningan tahun 2023. National Nursing Conference, 1(2), 55–68. https://doi.org/10.34305/nnc.v1i2.860
Purbosari, P. (2022). Strategi pemasaran program green laboratory menuju green hospital dengan metode segmenting, targeting, and positioning serta marketing mix di RSUP Dr. Sardjito, Yogyakarta. Jurnal ARSI (Administrasi Rumah Sakit Indonesia), 7(1), 2.
Sharma, A., Nishadham, V., Gupta, P., Gupta, G., Sharma, D., Goel, S., Pasricha, S., Kamboj, M., & Mehta, A. (2025). Evaluation of turnaround times of diagnostic biopsies: A metric of quality in surgical pathology. International Journal of Surgical Pathology, 33(2), 344–352. https://doi.org/10.1177/10668969241261561
Sutrisna, I. G. P. A. F., Prihatiningsih, D., & Subhaktiyasa, P. G. (2020). Analisis pemantapan mutu internal pemeriksaan trombosit di laboratorium klinik UPTD Puskesmas Abiansemal I. Jurnal Analis Laboratorium Medik, 5(2), 28–34. https://doi.org/10.51544/jalm.v5i2.2152
Swastini, I. G. A. A. P. (2021). Gambaran kolesterol total pada lansia di Puskesmas I Denpasar Selatan. Meditory: The Journal of Medical Laboratory, 9(2), 68–77. https://doi.org/10.33992/m.v9i2.1526
Wang, X., & Hu, L. (2020). Enzymatic strips for detection of serum total cholesterol with point-of-care testing (POCT) devices: Current status and future prospect. Journal of The Electrochemical Society, 167(3), 037535. https://doi.org/10.1149/1945-7111/ab64bb
Zhao, X., Qin, X., & Miao, Y. (2025). Reducing turnaround time for routine outpatient biochemical tests through Lean Six Sigma: A case study in China. Journal of Evaluation in Clinical Practice, 31(1), e14116. https://doi.org/10.1111/jep.14116
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2025 OBAT: Jurnal Riset Ilmu Farmasi dan Kesehatan

This work is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.