Perbedaan Hasil Pemeriksaan Kolesterol Total dalam Darah Menggunakan Alat Architect C4000 dan Alinity C1000

Authors

  • Retno Rusnaini Sekolah Tinggi Ilmu Kesehatan Wira Medika
  • Nyoman Sudarma Sekolah Tinggi Ilmu Kesehatan Wira Medika
  • Ni Luh Gede Puspita Yanti Sekolah Tinggi Ilmu Kesehatan Wira Medika

DOI:

https://doi.org/10.61132/obat.v3i6.1802

Keywords:

Alinity, Architect, Cholesterol, Enzymatic colorimetric, Hypercholesterolemia

Abstract

Total cholesterol measurement is essential for assessing cardiovascular risk, but consistency between laboratory instruments must be validated for reliable results. Architect C4000 and Alinity C1000 operate on enzymatic principles but differ in system design, reagents, and supporting technology. This study aimed to compare total cholesterol measurements between the two instruments using a non-parametric statistical approach (Mann-Whitney test). A comparative analytical design was conducted with 100 serum samples obtained via purposive sampling and divided for measurement with each device. Statistical analysis evaluated differences, and mean differences were calculated using the Hodges-Lehman method. Results showed mean cholesterol values of 182.98 mg/dL for Architect and 182.37 mg/dL for Alinity, with no significant difference (p=0.9942) and a median difference of 0.0 (95% CI: -9.0 to 9.0). Data distribution was nearly identical for both instruments in terms of mean and spread. Therefore, both methods demonstrate comparable validity in total cholesterol measurement, as they rely on similar principles and technologies, with reagents calibrated to the same standards.

Downloads

Download data is not yet available.

References

Aisya, R. W., & Dharmawati, L. (2021). Hubungan kebiasaan konsumsi makanan cepat saji dengan kejadian penyakit jantung koroner pada pasien rawat jalan di RSUD Dr. Moewardi. Jurnal Medika Indonesia, 2(2), 21–28.

Bishop, M. L. (2020). Clinical chemistry: Principles, techniques, and correlations, enhanced edition. Jones & Bartlett Learning.

Caeiro, A., Caeiro, S., Correia, S., & Canhoto, J. (2022). Induction of somatic embryogenesis in tamarillo (Solanum betaceum Cav.) involves increases in the endogenous auxin indole-3-acetic acid. Plants, 11(10), 1347. https://doi.org/10.3390/plants11101347

Desoky, E.-S. M., Mansour, E., Ali, M. M. A., Yasin, M. A. T., Abdul-Hamid, M. I. E., Rady, M. M., & Ali, E. F. (2021). Exogenously used 24-epibrassinolide promotes drought tolerance in maize hybrids by improving plant and water productivity in an arid environment. Plants, 10(2), 354. https://doi.org/10.3390/plants10020354

Douxfils, J., Morimont, L., Delvigne, A.-S., Devel, P., Masereel, B., Haguet, H., Bouvy, C., & Dogné, J.-M. (2020). Validation and standardization of the ETP-based activated protein C resistance test for the clinical investigation of steroid contraceptives in women: An unmet clinical and regulatory need. Clinical Chemistry and Laboratory Medicine, 58(2), 294–305. https://doi.org/10.1515/cclm-2019-0471

Fan, G., & Wang, Q. (2021). Quality control and quality assurance. In Clinical molecular diagnostics (pp. 97–113). Springer. https://doi.org/10.1007/978-981-16-1037-0_8

Juraijin, D. (2025). Quality assurance analysis internal analytical phase of urine microalbumin examination in patients hemodialisa. Journal Health Applied Science and Technology, 3(1), 25–30. https://doi.org/10.52523/jhast.v3i1.68

Kopacz, A., Kubicka-Russel, D., Liszewski, G., Bukowska, A., Samek, S., Malka, D., Łętowska, M., & Grabarczyk, P. (2024). Evaluation and experience from routine use of chemiluminescence assays for serological screening of blood and plasma donations on the Alinity s system and the Alinity i system, two new fully automated immunoassay systems in Poland. Practical Laboratory Medicine, 39, e00364. https://doi.org/10.1016/j.plabm.2024.e00364

Lapić, I., Šegulja, D., Dukić, K., Bogić, A., Lončar Vrančić, A., Komljenović, S., Šparakl, T., Grdiša Teodorović, K., Cigula Kurajica, V., & Baršić Lapić, I. (2022). Analytical validation of 39 clinical chemistry tests and 17 immunoassays on the Alinity analytical system. Scandinavian Journal of Clinical and Laboratory Investigation, 82(3), 199–209. https://doi.org/10.1080/00365513.2022.2056856

Liana, P., Jenica, A., Suciati, T., Rahmawati, E., Pariyana, P., & Umar, T. P. (2023). Comparison of liver function test results between Architect C8000 and COBAS C501 automatic chemistry analyzer. Archives of Razi Institute, 78(3), 1141.

Moonti, M. A., Sutandi, A., & Fitriani, N. D. (2023). Hubungan life style dengan kejadian hipertensi pada dewasa di Desa Jagara Kecamatan Darma Kabupaten Kuningan tahun 2023. National Nursing Conference, 1(2), 55–68. https://doi.org/10.34305/nnc.v1i2.860

Purbosari, P. (2022). Strategi pemasaran program green laboratory menuju green hospital dengan metode segmenting, targeting, and positioning serta marketing mix di RSUP Dr. Sardjito, Yogyakarta. Jurnal ARSI (Administrasi Rumah Sakit Indonesia), 7(1), 2.

Sharma, A., Nishadham, V., Gupta, P., Gupta, G., Sharma, D., Goel, S., Pasricha, S., Kamboj, M., & Mehta, A. (2025). Evaluation of turnaround times of diagnostic biopsies: A metric of quality in surgical pathology. International Journal of Surgical Pathology, 33(2), 344–352. https://doi.org/10.1177/10668969241261561

Sutrisna, I. G. P. A. F., Prihatiningsih, D., & Subhaktiyasa, P. G. (2020). Analisis pemantapan mutu internal pemeriksaan trombosit di laboratorium klinik UPTD Puskesmas Abiansemal I. Jurnal Analis Laboratorium Medik, 5(2), 28–34. https://doi.org/10.51544/jalm.v5i2.2152

Swastini, I. G. A. A. P. (2021). Gambaran kolesterol total pada lansia di Puskesmas I Denpasar Selatan. Meditory: The Journal of Medical Laboratory, 9(2), 68–77. https://doi.org/10.33992/m.v9i2.1526

Wang, X., & Hu, L. (2020). Enzymatic strips for detection of serum total cholesterol with point-of-care testing (POCT) devices: Current status and future prospect. Journal of The Electrochemical Society, 167(3), 037535. https://doi.org/10.1149/1945-7111/ab64bb

Zhao, X., Qin, X., & Miao, Y. (2025). Reducing turnaround time for routine outpatient biochemical tests through Lean Six Sigma: A case study in China. Journal of Evaluation in Clinical Practice, 31(1), e14116. https://doi.org/10.1111/jep.14116

Downloads

Published

2025-11-06

How to Cite

Retno Rusnaini, Nyoman Sudarma, & Ni Luh Gede Puspita Yanti. (2025). Perbedaan Hasil Pemeriksaan Kolesterol Total dalam Darah Menggunakan Alat Architect C4000 dan Alinity C1000. OBAT: Jurnal Riset Ilmu Farmasi Dan Kesehatan, 3(6), 01–13. https://doi.org/10.61132/obat.v3i6.1802

Similar Articles

You may also start an advanced similarity search for this article.